Recall Z-1100-2026 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

ICU Medical Inc.

Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG

DateJan 15, 2026 (initiated Dec 19, 2025)
StatusOpen, Classified
ReasonAffected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.
Root causeUnder Investigation by firm
ActionOn December 19. 2025, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were informed that affected lots may have a crack present on the syringe collar. Customers were instructed to discontinue use of affected lots, and to destroy products following their institution's process for destruction. If destruction is not immediately possible, affected product should be quarantined until disposal.
Quantity64290 total
DistributionUS Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO, NE, NJ, NY, OH, OK, TX, WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952516PRO-VENT, LINE DRAW, PULSATOR
Smiths Industries Medical Systems, Inc., Jul 14, 1995