Recall Z-1121-2026 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2

DateJan 16, 2026 (initiated Nov 28, 2025)
StatusOpen, Classified
ReasonCertain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 11/28/25 was sent to customers. Actions to be Taken by Customers 1. Immediately locate, isolate, and cease all use of the impacted product (see Attachment A for complete list of affected serial numbers). The product code and serial number can be found on the bottom of the infusion pump. 2. If you received this communication directly from Baxter, please acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will preven
Quantity585 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K251640SIGMA Spectrum Infusion Pump with Master Drug Library
Baxter Healthcare Corporation, Jul 28, 2025