ClearanceBrief
Recall Z-1059-2026 ยท OJG Neurological Tray

Medline Industries, LP

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: NEURO ANGIO PACK, Medline kit SKU DYNJ38935

DateJan 12, 2026 (initiated Nov 26, 2025)
StatusOpen, Classified
ReasonMedline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Root causeNonconforming Material/Component
ActionMedline issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue potential risk to the patient, and referred the consignee to the attached B Braun notice for further instructions. The B Braun notice provide "Interim Measures for Users" to provide for the continued use of the product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity12 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.