Recall Z-1042-2026 ยท LIT Catheter, Angioplasty, Peripheral, Transluminal

Boston Scientific Corporation

Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF H74939186251210; 2) 3.5mm x120mm, 150 cm, UPM/Model No. REF H74939186351210; Catheter, Angioplasty, Peripheral, Transluminal

DateJan 9, 2026 (initiated Dec 3, 2025)
StatusOpen, Classified
ReasonBoston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation difficulty during use in the peripheral vasculature. An internal investigation found that a manufacturing process variation resulted in catheters that may not properly inflate/deflate. This was limited to Coyote OTW batches manufactured from 12Sep2025 through 13Oct2025. The processing variation has since been addressed. No other material numbers or batches are impacted by this r
Root causeProcess change control
ActionBoston Scientific issued an Urgent Medical Device Removal notice to its consignees on 12/3/2025 by letter delivered via a courier service. The notice explained the issue, clinical impact to the patient, and requested the following actions: "Instructions: 1. Further distribution or use of any remaining devices (Attachment 1) affected by this removal should cease immediately. Do NOT use affected devices and remove those devices from your facility s inventory. Segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visibl
Quantity21 units
DistributionUS Nationwide distribution in the states of MA, WV, OH, CA, MN, NY, VA, IL TX, FL, IL.