Recall Z-1157-2026 ยท CAK Microfilter, Blood Transfusion

GVS TM, Inc

SQ40S Blood Transfusion Filter

DateJan 22, 2026 (initiated Nov 3, 2025)
StatusOpen, Classified
ReasonDue to products being released for distribution prior to completion of the required quality control release process the sterility assurance cannot be confirmed.
Root causeProcess control
ActionOn 11/03/2025, the firm sent via email a "GVS TM Inc Customer Statement" informing customers that GVS TM Inc discovered that some of the products in the affected lot may have been damaged during the transportation process. To avoid any potential risk this damage may pose to the downstream user, the firm is requesting return of the affected products. On 11/18/2025, the firm sent via email a "GVS TM Inc Product Recall Statements" informing customers that the products were distributed before completion of the required quality control release process, including final review of sterilization docu
Quantity2,720 filters
DistributionU.S.: ID, NV, SC, MO, TX, CO, MD, CA, VA, DC, PA, IL, OH, AZ, FL, OK, MN, KY, OR, NE, NC, LA, MS, MI, NY, GA, AL, MA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K960993PALL SQ40S BLOOD FILTER (SQ40S/SQ40ST)
Pall Biomedical Products Co., Apr 22, 1996