ClearanceBrief
Recall Z-1137-2026 ยท FSA Lift, Patient, Non-Ac-Powered

Community Products, LLC

Rifton Equipment, TRAM. Model/Catalog Number: K310. The TRAM is a powered lift device for transferring clients in a seated position or assisting in a sit-to-stand transfer and gait training.

DateJan 22, 2026 (initiated Dec 18, 2025)
StatusOpen, Classified
ReasonPotential for fraying on the body support strap.
Root causeProcess change control
ActionCommunity Products notified consignees via email on about 12/18/2025. Consignees were instructed to notify all customers, if further distributed, of the recall, inspect all affected devices for visible fraying, and cease use if any is observed. Consignees were requested to provide contact information for the individual responsible for performing the replacement. If devices were further distributed, consignees were advised to manage all customer outreach, inspection and replacement. The other option was to provide Community Products with customer contact information, and Rifton will manage the
Quantity3,238 units
DistributionWorldwide distribution - US Nationwide and the countries of UAE, Australia, Canada, EU, Israel, Hong Kong, Japan, Korea, New Zealand, Russia, Saudi Arabia, and UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.