Recall Z-1123-2026 ยท PTM Syringe, Balloon Inflation, Exempt

Merit Medical Systems, Inc.

Merit Medical Inflation Device.basixCOMPAK.30 atm/bar.20 mL Reference Numbers: ENDO-AN2030/B IN4130/CNK IN4130/H IN4130/JPH IN4130/K IN4130/KRK IN4130/L IN4130/T IN4230/JPH IN4230/K IN4330/CNK IN4330/K IN4352/K IN4430/K IN4530/CNK IN4530/K IN4530/T IN4802/K IN4802/T IN4852/K IN4901/A IN4902/A IN4903/A IN4904/A IN4905/A IN4906/A IN4907/A IN4908 IN4909 IN4910 IN4911/A IN4912 IN4913/A IN4914/A IN491

DateJan 16, 2026 (initiated Dec 4, 2025)
StatusOpen, Classified
ReasonInflation device handle may detach from the syringe during procedure.
Root causeProcess control
ActionOn December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the produc
Quantity4,163,123 units
DistributionWorldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY, St. Thomas , Puerto Rico and the