Recall Z-1094-2026 · FRN Pump, Infusion

ICU Medical, Inc.

Plum Duo Infusion System, List Number 40002-04-01

DateJan 15, 2026 (initiated Dec 15, 2025)
StatusOpen, Classified
ReasonICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs.
Root causeNonconforming Material/Component
ActionAn URGENT: MEDICAL DEVICE CORRECTION dated 12/15/25 was sent to customers. Actions for Users: 1. Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately informed of this notice. If the devices are used at another location, please ensure this communication is delivered to that location. 2. If the primary speaker fails, obtain another pump to continue the infusion. Remove the affected pump from clinical use until the speaker can be replaced. 3. Complete and return the attached Customer Response Form to icumedmarketactionstea
Quantity423 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223607Plum Duo™ Infusion System
Icu Medical, Inc., Aug 24, 2023