Recall Z-1299-2026 ยท KPR System, X-Ray, Stationary

FUJIFILM Healthcare Americas Corporation

FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

DateFeb 5, 2026 (initiated Jan 9, 2026)
StatusOpen, Classified
ReasonIt was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Root causeProcess control
ActionOn January 9, 2026 URGENT: MEDICAL DEVICE RECALL letters were emailed to customers. Actions to be taken: Please pay attention to the following points until the work is completed. " For safety reasons, please turn off the breaker of the distribution board to which the power cord of the unit is connected after turning off the power of the unit according to the operation manual. " Do not forget to turn off the breaker of the distribution board until the replacement of the contactor is completed. 1) Please post this notification on or near the device 2) Please complete and return the Field Action
Quantity16 units
DistributionUS Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152294FDR Visionary Suite
Shimadzu Corporation, Jul 15, 2016