Recall Z-1237-2026 ยท KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Tornier, Inc

TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318

DateFeb 2, 2026 (initiated Dec 19, 2025)
StatusOpen, Classified
ReasonA specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.
Root causeLabeling mix-ups
ActionOn December 19, 2025, the firm began notifying affected customers via Urgent Field Safety Notice letters. Customers were instructed to immediately check their internal inventory for affected product. Firm stated that return of product is not necessary because the issue is detectable pre-operatively. The physical product itself is correctly marked, and after unpacking, should only be assigned to the correct place in the screw caddy. The notification is being provided for awareness only.
Quantity427
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161742Aequalis PerFORM Reversed, Aequalis PerFORM+ Reversed Glenoid
Tornier, Inc., Nov 15, 2016