Recall Z-1031-2026 ยท IQA System, Environmental Control, Powered

Securitas Healthcare LLC

Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3

DateJan 8, 2026 (initiated Dec 8, 2025)
StatusOpen, Classified
ReasonSecuritas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.
Root causeUnder Investigation by firm
ActionA notification letter dated 12/1/25 was sent to customers. URGENT: SECURITAS HEALTHCARE RECALLS CSK200-900 AND CSK200-900MR SERIES CALL STATIONS DUE TO RISK THAT LOW BATTERY ALERTS WILL NOT BE RECEIVED BY THE ARIAL WIRELESS EMERGENCY CALL SYSTEM BEFORE DEPLETION OF BATTERY. Customer Required Actions: 1. Reduce the Missing Device Timeout to 125 Minutes for all Call Stations. Clients should reduce the Missing Device Timeout to 125 minutes so that a Missing Device Alert will be triggered if a Call Station fails to send a supervision message to the Arial Emergency and Nurse Call System wi
Quantity5627 units
DistributionWorldwide distribution -US Nationwide and the country of Canada.
FDA recordOpen on fda.gov