Recall Z-1214-2026 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America

Upgrade to MR 7700. Product Code (REF): 782130. MR systems with SW version R11.1 and R12.1.

DateJan 29, 2026 (initiated Dec 3, 2025)
StatusOpen, Classified
ReasonThe potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Root causeSoftware design
ActionPhilips began notification of consignees on about 12/03/2025 via mailed Urgent Medical Device Correction letter, with a corrected notification sent on about 12/18/2025. Consignees were instructed to identify if any affected units were on hand, that the systems could continue to be use in accordance with their intended use. They were asked to provide the notification to all users of the device for awareness, maintain the Urgent Medical Device Correction letter with affected systems until the software upgrade is installed, and complete and return the provided response form. A Philips represe
Quantity6 units
DistributionWorldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, Unite
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213516Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei
Philips Medical Systems Nederland B.V., Mar 3, 2022
K223442MR 5300 and MR 7700 R11 MR Systems
Philips Medical Systems Nederlands B.V., Dec 23, 2022
K230972Ingenia Elition and MR 7700 MR Systems
Philips Medical Systems Nederland B.V., Apr 28, 2023