Recall Z-1018-2026 ยท LNH System, Nuclear Magnetic Resonance Imaging

Canon Medical System, USA, INC.

VANTAGE TITAN 3T Model MRT-3010/MEXL-3010

DateJan 13, 2026 (initiated Dec 19, 2025)
StatusOpen, Classified
ReasonThere is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
Root causeUnder Investigation by firm
ActionOn 12/19/2025, the firm sent via Adobe DocuSign an "Urgent: Medical Device Correction" letter to customers informing them that It has been discovered that in the superconducting magnet used in the applicable system, if there is slow leakage, over a long period of time, of refrigerant gas from the superconducting magnet, air ingress into the helium vessel via the leak point may result in ice forming. If this is left unchecked, the worst-case scenario is that the helium gas will not be able to escape through the designed exhaust path, and the helium tank could eventually rupture, releasing heliu
Quantity22 systems
DistributionU.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191128Vantage Titan 3T, MRT-3010, V2.5
Canon Medical Systems Corporation, Jun 26, 2019