Recall Z-1048-2026 ยท FRN Pump, Infusion

Fresenius Kabi USA, LLC

Ivenix Infusion System (IIS), LVP Software: LVP-SW-0005.

DateJan 12, 2026 (initiated Nov 21, 2025)
StatusOpen, Classified
ReasonEmphasizing instructions for LVP duration programming located in the IFU.
Root causeUse error
ActionAn URGENT MEDICAL DEVICE CORRECTION notification dated 11/21/25 was sent to consignees notifying them of this safety alert. The notification reiterates instructions for administering infusions included in the device's IFU. Fresenius Kabi asks consignees to disseminate the notification to all users and any facilities where devices may have been further distributed for their awareness. Consignees with any questions are to contact Customer Support by email at Ivenix_support@fresenius-kabi.com or by phone at 1-855-354-6387.
Quantity30 units
DistributionUS Nationwide distribution in the states of CA, CO, FL, GA, ID, IL, MD, MI, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183311Ivenix Infusion System (IIS)
Ivenix, Inc., Jun 7, 2019