Recall Z-1009-2026 ยท JZO System, Test, Thyroid Autoantibody

Roche Diagnostics Operations, Inc.

The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and electrochemiluminescence detection. Patient serum sample is incubated with pretreatment reagent PT1 and the reconstituted PT2 containing a pre-formed immunocomplex of solubilized porcine TSH receptor (pTSHR) and biotinylated anti-pTSHR mouse monoclonal antibody.

DateJan 2, 2026 (initiated Nov 21, 2025)
StatusOpen, Classified
ReasonIssues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.
Root causeProcess control
ActionFirm notified customers on November 21, 2025 via Urgent Medical Device Correction notices. Customers using the cobas e 402/801 and cobas e411 were instructed to immediately discontinue using the impacted lots outlined in the UMDC and switch to the new unaffected lot numbers listed in the communication.
Quantity1300 cassettes
DistributionUS Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, ME, MI, MN, NC, NJ, NM, NY, OH, PA, PR, TN, TX, UT, WA, WV.