Recall Z-1007-2026 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Ultrasound, LLC

Philips EPIQ Ultrasound Systems (various models) used in conjunction with the X5-1c transducer REF: , , 795200, 795201, 795231, 795232, 795234 REF: 795117, 795122

DateDec 31, 2025 (initiated Jul 14, 2022)
StatusCompleted
ReasonRetroactive: Ultrasound may experience an unexpected automatic reboot, resulting in damage to transducer.
Root causeSoftware Design Change
ActionOn July 18, 2022, Phillips issued a product notification to affected consignees. Philips ask consignees to take the following actions: 1. If the X5-1c transducer is connected to the right-most port (when viewing from the front), this problem will not occur. Please ensure your X5-1c is connected to the right-most connector until the software update (FCO79500561) has been installed. 2. Please retain this letter with your product documentation until the software update (FCO79500561) has been implemented on your system. 3. Please circulate this notice to all users of this device so that they a
Quantity1,721 systems
DistributionWorldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Argentina, Australia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K231190EPIQ Series Diagnostic Ultrasound System
Philips Ultrasound, LLC, May 12, 2023