Recall Z-0923-2026 ยท IZF System, X-Ray, Tomographic

AGFA Healthcare Corp.

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

DateDec 22, 2025 (initiated Nov 18, 2025)
StatusOpen, Classified
ReasonIt is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
Root causeRadiation Control for Health and Safety Act
ActionAGFA NV notified Customers via a letter titled "IMPORTANT PRODUCT INFORMATION LETTER / DR 800", included Product Name, Symptom, Cause, Actions to be taken customer and manufacturers respectively: "Always enable ABS for pulsed fluoro exams. Actions being taken by the manufacturer: Agfa/local business partner will contact you to install a new generator software."
Quantity35
DistributionU.S. Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183275DR 800 with Tomosynthesis
Agfa N.V., Feb 1, 2019