ClearanceBrief
Recall Z-0915-2026 ยท BSZ Gas-Machine, Anesthesia

GE Medical Systems China Co., Ltd.

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

DateDec 18, 2025 (initiated Nov 14, 2025)
StatusOpen, Classified
ReasonPotential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may oc
Root causeComponent change control
ActionGE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 11/14/2025 via letter using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken by Customer/User: "Pending correction by GE HealthCare, you can continue to use the Anesthesia machine with affected power management boards by following the instructions below: 1. Always ensure the device has a secure connection to an AC mains power source. 2. If there is a loss of AC mains power to the system leading to an unexpected system shutdown: - Promptly i
Quantity203 units
DistributionWorldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, H
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151570Carestation 620/650/650C
Datex-Ohmeda, Inc., Jan 21, 2016
K213867Carestation 750/750c
Datex-Ohmeda, Inc., Mar 28, 2023