Recall Z-1093-2026 ยท LEA Preservative, Cytological

Hologic, Inc

ThinPrep CytoLyt Solution. Buffered Cell Wash Solution.

DateJan 15, 2026 (initiated Dec 12, 2025)
StatusOpen, Classified
ReasonFungal contamination of affected lot with Parengyodontium album.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE RECALL NOTIFICATION was mailed to consignees on 12/12/25. The notification informs consignees that Hologic will provide replacement bottles of ThinPrep CytoLyt Solution to affected customers from unaffected lots based on retained sample testing. Consignees are asked to inspect their inventory for affected product and discontinue use immediately. Affected product is to be quarantined and destroyed per routine laboratory disposal procedures. Consignees are asked to complete the Customer Response Form using the provided QR Code and Hologic will coordinate replacement. Con
Quantity3,728 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.