Recall Z-0597-2026 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips North America

Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM

DateDec 9, 2025 (initiated Oct 27, 2025)
StatusOpen, Classified
ReasonBetween July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
Root causeRadiation Control for Health and Safety Act
ActionPhilips will send a customer notification to the users of the affected systems to alert them of the issue and provide the actions that should be taken to prevent risks to patients. Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit their site and install the SSD spacer to resolve the issue.
Quantity110
DistributionU.S.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200917Azurion R2.1
Philips Medical Systems Nederland B.V., May 1, 2020