Recall Z-1341-2026 ยท JCX Apparatus, Suction, Ward Use, Portable, Ac-Powered

Laerdal Medical Corporation

Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;

DateFeb 11, 2026 (initiated Jan 13, 2026)
StatusOpen, Classified
Reasonunits manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.
Root causeNonconforming Material/Component
ActionOn January 13, 2026, Medical Device Recall letters were sent to customers. Actions to be taken: 1. Please review this notice carefully and provide acknowledgment that you have received and understood the contents. If you received this notice from a distributor or other third party, please send your acknowledgment directly to Laerdal using the contact information provided below. You must provide the case number in the footer of this notice in your response. 2. Ensure that this notice is shared with all relevant personnel involved in the storage, maintenance, charging, and use of the LCSU 4 RTC
Quantity1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)
DistributionWorldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120065PLAXTRON SUCTION UNIT, MODEL 88AA51/88AA61 SERIES
Plaxtron Industrial (M) Sdn. Bhd., Apr 3, 2012