Recall Z-0961-2026 · JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories

Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20

DateDec 18, 2025 (initiated Nov 13, 2025)
StatusOpen, Classified
ReasonAbbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the recall is potential performance issues found in the Alinity ci-series System software versions 3.6.1. and lower that could lead to erroneous results for multiple analytes. The issue was identified by Abbott during the internal testing of complaint investigations.
Root causeSoftware design
ActionOn 11/13/2025, "Urgent Field Safety Notice Product Correction" notices were sent to customers who were asked to take the following actions: 1) Perform the Weekly maintenance procedures (i-series) 2630 Manual Wash Cup Cleaning (i-series) per Section 9 of the Alinity ci-series Operations Manual. 2) Field Service Representative will contact you to schedule installation of software version 3.7.0. 3) It is recommended to perform the Pipettor diagnostic procedures (c-series) 4109 Probe Alignment Test (c-series) daily for R1 and R2 pipettor probes per Section 9 of the Alinity ci-series Operations Man
Quantity11653
DistributionWorldwide distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NH, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR; and OUS (foreign) to countries of: Al
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170316ALINITY C GLUCOSE REAGENT KIT, ALINITY C SYSTEM
Abbott Laboratories, Oct 19, 2017
K170317Alinity i Total ß-hCG Reagent Kit, Alinity i System
Abbott Laboratories, Oct 23, 2017