Recall Z-0996-2026 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Medtronic Neuromodulation

Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET EU -UK CT900C PROG CT900C CLINICIAN TABLET UK CT900D PROG CT900D CLINICIAN TABLET GLOBAL CT900E PROG CT900E CLINICIAN TABLET GLBL SMSNG CT900F PROG CT900F CLINICIAN TABLET GLBL SMSNG

DateDec 23, 2025 (initiated Nov 20, 2025)
StatusOpen, Classified
ReasonSoftware issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.
Root causeSoftware design
ActionOn November 20, 2025, the firm notified customers via Urgent Medical Device Safety Notification letters. Medtronic has developed a new version of the A71100 CP App, v1.0.4489 or later, that contains a fix for this issue by ensuring the ability to clear a Device Reset message by pressing OK if the message appears. Customers should download v1.0.4489 or any later version at your earliest convenience.
Quantity7667 devices
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, French Pol