Recall Z-0949-2026 ยท MIB Elastomer, Silicone Block
DSAART LLC
Brand Name: AART Calf Implant Product Name: Calf Implant Model/Catalog Number: 502-103 - Calf Implant style 1 size 3 UDI-DI code: B490502103 502-104 - Calf Implant style 1 size 4 UDI-DI code: B490502104 502-105 - Calf Implant style 1 size 5 UDI-DI code: B490502105 502-106 - Calf Implant style 1 size 6 UDI-DI code: B490502106 502-201 - Calf Implant style 2 size 1 UDI-DI code: B490502201 502-30
| Date | Dec 18, 2025 (initiated Sep 26, 2025) |
| Status | Open, Classified |
| Reason | Manufactured products deviated from established specifications and were distributed without obtaining the required FDA premarket approvals or clearances. |
| Root cause | No Marketing Application |
| Action | On 09/26/2025, the firm emailed a notice to an affected customer informing them that due to identified testing issues, the firm was requesting return of gluteal implants. On 11/17/2025, the firm send an "URGENT Medical Device removal of AART Silicone Implants" Letter to customer informing them that Alpha Aesthetics is voluntarily removing products as they conduct an analysis to determine the regulatory status of the devices. Customers are instructed to: 1. Either return the affected products that have not been implanted to Alpha Aesthetics at the company's expense, or discard affected produc |
| Quantity | 35 implants |
| Distribution | U.S.: AR, AZ, CA, CO, CT, DC, FL, GA, KY, IL, MA, MD, MI, MN, MS, NE, NJ, NY, OH, OK, PA, TX, UT, VA, WA, WI and Puerto Rico. O.U.S.: N/A |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K022511 | AART CALF IMPLANT Aesthetic and Reconstructive Technologies, Inc., Aug 26, 2002 |