Recall Z-0885-2026 ยท FRN Pump, Infusion

Fresenius Kabi USA, LLC

Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.

DateDec 16, 2025 (initiated Nov 14, 2025)
StatusOpen, Classified
ReasonSoftware version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.
Root causeSoftware design
ActionConsignees were emailed an "URGENT - Medical Device Field Correction" notice dated 11/14/25. The notice instructs consignees to review and become familiar with the provided notification's risk mitigation actions; the notice is to be forwarded if affected units were further distributed and users are to be notified. Consignees are to facilitate the installation of the new Ivenix Infusion Management System (version 5.2.2) and Large Volume Pump (5.10.2) software versions utilizing the provided instructions. Consignees are to note that the LVP will not be available for use during the update. Comple
Quantity30 units
DistributionUS (Domestic) distribution: CA, CO, FL, GA, ID, IL, MD, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA & WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183311Ivenix Infusion System (IIS)
Ivenix, Inc., Jun 7, 2019