Recall Z-0933-2026 ยท CAD Vaporizer, Anesthesia, Non-Heated

Draeger, Inc.

Draeger Vapor 2000. Unheated, calibrated anesthetic vaporizer.

DateDec 23, 2025 (initiated Nov 24, 2025)
StatusOpen, Classified
ReasonA certain component of affected devices was not delivered within specification and contained impurities.
Root causeUnder Investigation by firm
ActionConsignees were mailed an Urgent Medical Device Recall notification dated November 2025. The notification recommends that consignees do not use affected vaporizers and return the provided reply card to confirm this. A Draeger representative will contact consignees to arrange replacement of affected vaporizers. If affected vaporizers have been placed into clinical use, Draeger asks that consignees clean the breathing system of the anesthesia device according to the IFU and use of inspiratory filters is recommended. Consignees are to share the recall notification with all users of affected devic
Quantity431 units
DistributionWorldwide - US Nationwide distribution in the states of AK, CA, FL, IA, KS, MD, ME, MN, NY, OH, PA, RI, SD, TN, TX, UT, W and the countries of Argentina, Australia, Belgium, Brazil, Cambodia, China, Columbia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Hungary, Indonesia, Isr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K971923DRAGER-VAPOR 2000
Draeger Medical, Inc., Sep 16, 1997