Recall Z-0935-2026 ยท LLZ System, Image Processing, Radiological

GE Healthcare GmbH

GE HealthCare ViewPoint 6, Catalog Number H47601AA; Radiological image processing system

DateDec 16, 2025 (initiated Nov 3, 2025)
StatusOpen, Classified
ReasonGE HealthCare has become aware that the Patient History tab in ViewPoint 6 versions 6.15.3 and 6.15.4 ultrasound software may not include complete patient medical history information.
Root causeDevice Design
ActionGE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/3/2025 via letter distributed using a traceable means. The notice explained the issue, safety issue/risk, and requested the following: Actions to be taken by Customer/User: "Pending correction by GE HealthCare, you can continue to use your device by following the instructions below: When creating or opening an exam on ViewPoint 6 version 6.15.3 or ViewPoint 6 version 6.15.4, if using the patient history tab, do not rely on the information. Always confirm that essential patient history data is correc
Quantity1221 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K241300ViewPoint 6
Ge Healthcare GmbH, Jul 2, 2024