Recall Z-0955-2026 ยท GBX Catheter, Irrigation
Rocket Medical Plc
Brand Name: Rocket Product Name: Rocket 20Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54549-20-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle dev
| Date | Dec 18, 2025 (initiated Nov 7, 2025) |
| Status | Open, Classified |
| Reason | IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use. |
| Root cause | Labeling design |
| Action | On November 12, 2025, Urgent Medical Device Correction Letters were sent to customers. Actions to be taken: 1. Ensure a copy of this letter and the updated IFU are available to all users or potential users of this device. 2. Complete the FSN acknowledgement form and return via the email provided. |
| Quantity | 246 units |
| Distribution | Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, S |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.