ClearanceBrief
Recall Z-0950-2026 ยท GBX Catheter, Irrigation

Rocket Medical Plc

Brand Name: Rocket Product Name: Rocket 12Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54544-12-PK Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A

DateDec 18, 2025 (initiated Nov 7, 2025)
StatusOpen, Classified
ReasonIFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Root causeLabeling design
ActionOn November 12, 2025, Urgent Medical Device Correction Letters were sent to customers. Actions to be taken: 1. Ensure a copy of this letter and the updated IFU are available to all users or potential users of this device. 2. Complete the FSN acknowledgement form and return via the email provided.
Quantity635 units
DistributionDomestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.