Recall Z-0917-2026 · DSP System, Balloon, Intra-Aortic And Control

Datascope Corp.

Cardiosave Rescue. Intra-Aortic Balloon Pump system.

DateDec 11, 2025 (initiated Nov 4, 2025)
StatusOpen, Classified
ReasonThe IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.
Root causeComponent design/selection
ActionAn URGENT MEDICAL DEVICE CORRECTION NOTICE, dated October 2025, was sent via FEDEX to consignees. The notification notifies consignees of four addendum updates: 1) NVRAM Preventive Maintenance (New FSCA - reflected in this recall entry), 2) Vibration and Shock Table Specifications (New FSCA), 3) Helium O-Ring Preventive Maintenance (FSCA 2249723-01/24/2023-004-C), and 4) Battery Runtime Specifications (FSCA 2249723-09/10/2021-001-R). The notification instructs consignees regarding the NVRAM Preventive Maintenance to review the provided IFU addendum, update local preventive maintenance procedur
Quantity47 units
DistributionDomestic: Nationwide Distribution; Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d Ivoire, Croatia, Cyprus, Cze