ClearanceBrief
Recall Z-0593-2026 ยท BZA Connector, Airway (Extension)

Draeger, Inc.

ErgoStar CM 60, Model/Catalog Number: MP01860, Airway connector for conduction of respiratory gases between the breathing system and the patient interface. (e.g. endotracheal tube mask) of a mechanically ventilated patient.

DateDec 8, 2025 (initiated Nov 10, 2025)
StatusOpen, Classified
ReasonMultiple complaints were reported in which cracks formed in the hose of specific catheter mounts.
Root causeUnder Investigation by firm
ActionOn November 10, 2025, Urgent Medical Device Recall letters were sent to customers. Necessary Action: Identify if you have any affected product. Do not use any unused ErgoStar CM 40/CM 45/CM 55/CM 60 in your stock. Unused stock should be removed and returned to Draeger. Draeger recommends the alternative product MP01850 ErgoStar CM 50 as a suitable replacement. Please ensure that all potential users in your facility are made aware of this Urgent Medical Device Recall Notice and the information contained herein. If you have forwarded affected products to third parties, please also forward this U
Quantity2050 units
DistributionWorldwide distribution. US states of CA, FL, IL, NY, OH, OK, PA, SC, TX, and WI; Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Czech Republic, Ecuador, Finland, France, Germany, Hungary, Indonesia, Italy, Japan, Mexico, Netherlands, Norway, Pakistan, Poland, Portugal,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.