Recall Z-0726-2026 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Medtronic, Inc.

Aurora EV-ICD and Clinical EV-ICD, single chamber, extravascular implantable cardioverter defibrillators (ICD), Product Number DVEA3E4

DateNov 28, 2025 (initiated Oct 29, 2025)
StatusOpen, Classified
ReasonThere is a potential for delayed time to high-voltage (HV) therapy should a rare sequence of events occur.
Root causeSoftware design
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated October 2025 was sent to customers. PATIENT MANAGEMENT RECOMMENDATIONS FOR PREVIOUSLY DISTRIBUTED DEVICES: Medtronic acknowledges that each patient requires unique clinical considerations. Based on internal investigation and consultation with our Independent Physician Quality Panel, Medtronic provides the following guidance: "Prophylactic device replacement is NOT recommended. "Schedule an in-clinic SmartSync interrogation to receive the software update, with the intent for all patients to receive theupdate within the next 3-6 mont
Quantity6591 units
DistributionWorldwide - US Nationwide distribution.