ClearanceBrief
Recall Z-0450-2026 ยท OHD Gynecological Laparoscopic Kit

Medline Industries, LP

Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE

DateNov 13, 2025 (initiated Oct 10, 2025)
StatusOpen, Classified
ReasonFirm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Root causeNonconforming Material/Component
ActionSafety notice was provided to customers on 10/10/2025. Follow-up "URGENT MEDICAL DEVICE RECALL" notification dated 10/21/2025 was sent out on 10/21/2025. Notification was through email and first-class mail. Customers were instructed to Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product and destroy affected kits on hand. No products are to be returned to Medline. Please use the link and the information below to complete your response form. Please list the quantity of affected product you
Quantity96,406 total units
DistributionInternational distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.