ClearanceBrief
Recall Z-0644-2026 ยท FMG Stopcock, I.V. Set

B Braun Medical Inc

IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. US1295 ULTRASITE IV ADMIN SET 60DROP; Catalog Number: 352394.

DateNov 26, 2025 (initiated Oct 29, 2025)
StatusOpen, Classified
ReasonPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root causeProcess control
ActionB Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administrati
Quantity1,250 units
DistributionWorldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022895ELCAM STOPCOCKS AND MANIFOLDS
Medical Device Consultants, Inc., Oct 18, 2002
K032393INDUCTION AND SAMPLING MANIFOLD (OR STOPCOCK)
Elcam Medical Acal, Sep 12, 2003