ClearanceBrief
Recall Z-1141-2026 ยท HGM System, Monitoring, Perinatal

Edan Diagnostics

Brand Name: EDAN Product Name: Fetal & Maternal Monitor Model/Catalog Number: F6, F9, F6 Express, F9 Express Product Description: The devices are bedside fetal and maternal monitors, which are used to monitor the physiological parameters of pregnant women including the fetus from 28 weeks gestation. The devices may be used antepartum as well as during labor and delivery. Component: No

DateJan 22, 2026 (initiated Oct 24, 2025)
StatusOpen, Classified
ReasonAs a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
Root causeDevice Design
ActionOn 10/24/2025, the firm emailed notifying customers of a product recall. On 11/06/2025, the firm emailed an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them of labeling issues involving cybersecurity which arose from a communication from the U.S. Food & Drug Administration. -First issue involves an inappropriate publicly routable IP address that was used as an example in the user manual for demonstrating user-configurable IP settings and Second issue involves an inappropriate publicly IP address that was used as the factory default configuration. Customer are instructe
Quantity3518
DistributionWorldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150901Fetal & Maternal Monitor, models F6, F6 Express, F9, F9 Express
Edan Instruments, Inc., Jul 1, 2015
K173042Fetal & Maternal Monitor: Models F6, F6 Express, F9, F9 Express; Fetal Monitor: Models F2, F3
Edan Instruments, Inc., Aug 30, 2018