Recall Z-1217-2026 ยท FRN Pump, Infusion

CareFusion 303, Inc.

BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fl

DateJan 29, 2026 (initiated Nov 6, 2025)
StatusOpen, Classified
ReasonDue to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network
Root causeUnder Investigation by firm
ActionOn 11/06/2025, the firm sent via mail and email an "URGENT MEDICAL DEVICE PRODUCT CORRECTION" Letter informing customers that during an electronic medical record interoperability workflow, a BD Alaris" PCU (Point of Care Unit) connected to an affected BD Alaris" Pump Module may disconnect from the hospital network. The connectivity issue relates to an unrecognized date and time ( datetime ) stored within the affected BD Alaris" Pump Module software. The software that was loaded onto the affected pump modules has a datetime that falls within the Daylight Savings Time (DST) adjustment. The Alari
Quantity189 devices
DistributionU.S. Nationwide distribution in the states of AZ, IN, OH, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211218BD Alaris System with Guardrails Suite MX v12.1.2
Carefusion 303, Inc., Jul 21, 2023