Recall Z-0723-2026 ยท LLZ System, Image Processing, Radiological

INTELERAD MEDICAL SYSTEMS INCORPORATED

IntelePACS - InteleConnect / TechPortal

DateNov 26, 2025 (initiated Oct 17, 2025)
StatusOpen, Classified
ReasonSoftware application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.
Root causeSoftware Design Change
ActionOn 10/17/2025, correction notices were emailed to customers who were asked to do the following: 1. Schedule Feature Release Upgrade (FRU) update immediately with the firm's Customer Success Manager (CSM). 2. Interim Mitigation: Until the update is applied, please advise all staff to verify the 'Relevant Clinical Info' field after any Case Editor modifications and consider temporarily documenting critical clinical information in alternative fields as backup. 3. complete and return the Acknowledgement Form via email to regulatory-affairs@intelerad.com. For more information or to ask any questi
Quantity19
DistributionUS Nationwide distribution in the states of AZ, CT, AL, IL, WA, MD, NJ,CA, TN, IN, FL, NC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192176IntelePACS
Intelerad Medical Systems Incorporated, Apr 2, 2020