ClearanceBrief
Recall Z-0636-2026 ยท FPA Set, Administration, Intravascular

B Braun Medical Inc

Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT SET W/ 6 B/C VALVES, 17 IN; Catalog Number: 490238.

DateNov 26, 2025 (initiated Oct 29, 2025)
StatusOpen, Classified
ReasonPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Root causeProcess control
ActionB Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administrati
Quantity11,800 units
DistributionWorldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K943770LOW VOLUME MULTIPORT
B. Braun of America, Inc., Apr 10, 1995