Recall Z-0500-2026 ยท JOZ System, Automated Platelet Aggregation

Accriva Diagnostics, Inc.

VerifyNow PRUTest Platelet Reactivity Test REF 85225 UDI-DI code: 10711234150078 The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade.

DateNov 13, 2025 (initiated Oct 21, 2025)
StatusOpen, Classified
ReasonDue to an device without a premarket clearance being incorrectly package and distributed.
Root causeUnder Investigation by firm
ActionOn 10/21/2025, the firm sent via email an "Urgent Medical Device Recall" Letter to customers informing them that Accriva Diagnostics, Inc. has identified that a small number of VerifyNow PRUTest Platelet Reactivity Test (85225) cartridge pouches from lot WP1559-01 may contain VerifyNow P2Y12 Test cartridges. VerifyNow P2Y12 Test cartridges are not cleared by the US-FDA for use in the United States. Customers are instructed to: 1. Please forward this communication to those within your organization who need to become aware of this matter and verify they understand the procedure and can complete
Quantity125 boxes (3,125 tests)
DistributionWorldwide - U.S. Nationwide distribution in the states of AK, CA, FL, IA, IL, MD, MI, NY, SD, and TX. The countries of Hong Kong, Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141427ACCUMETRICS,INC. VERIFYNOW PRUTEST, ACCUMETRICS, INC. VERIFYNOW PRUTEST
Accumetrics, Inc., Jul 24, 2015