Recall Z-0523-2026 ยท MLS Labor And Delivery Kit
Medline Industries, LP
MEDLINE convenience kits labeled as: 1) MAJOR VAGINAL HARPER PACK-LF, REF DYNJ0161768G; 2) ACH MAJOR VAGINAL PACK, REF DYNJ67757B.
| Date | Nov 14, 2025 (initiated Sep 30, 2025) |
| Status | Open, Classified |
| Reason | Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released. |
| Root cause | Under Investigation by firm |
| Action | Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2025 and 10/6/2025 via email and mail. The notice explained the issue with the component, potential risk, and requested the following actions: 1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information found in the notice to complete the response form. Please list the quantity of affected product in inventory on the form. Even if no affected product is in inventory, pleas |
| Quantity | 424 kits |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.