Medline Industries, LP

MEDLINE convenience kits labeled as: 1) GENERAL LAPAROSCOPY, REF CDS930027W; 2) GENERAL MAJOR CDS, REF CDS980245X; 3) BASIC GENERAL CDS, REF CDS980566F; 4) TOTAL KNEE CDS, REF CDS982823R; 5) BASIC CDS, REF CDS982936U; 6) UPPER EXTREMITY CDS, REF CDS983907M; 7) MAJOR ABDOMINAL CDS, REF CDS983908Q; 8) LOWER EXTREMITY CDS, REF CDS983909Q; 9) GENERAL SURGERY CDS, REF CDS983914M; 10) GENERAL

DateNov 14, 2025 (initiated Sep 30, 2025)
StatusOpen, Classified
ReasonMedline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
Root causeUnder Investigation by firm
ActionMedline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2025 and 10/6/2025 via email and mail. The notice explained the issue with the component, potential risk, and requested the following actions: 1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information found in the notice to complete the response form. Please list the quantity of affected product in inventory on the form. Even if no affected product is in inventory, pleas
Quantity20941 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.