ClearanceBrief
Recall Z-0492-2026 ยท DHB System, Test, Radioallergosorbent (Rast) Immunological

Siemens Healthcare Diagnostics, Inc.

3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;

DateNov 13, 2025 (initiated Oct 6, 2025)
StatusOpen, Classified
ReasonThe barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Root causeUnder Investigation by firm
ActionOn October 6, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions Perform the instructions provided below: " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " Discontinue use of and discard IMMULITE 2000 Oak Mix Specific Allergen Lot 627. " Review your inventory of the product to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. " Complete and return the Field Corre
Quantity2802 units (OUS only)
DistributionWorldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101572IMMULITE 2000 3GALLERGY SPECIFIC IGE ASSAY KIT
Siemens Healthcare Diagnostics, Aug 1, 2011