Recall Z-0883-2026 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Ultrasound, LLC

Lumify Diagnostic Ultrasound System models : 795005, 795161, 989605449841,

DateDec 4, 2025 (initiated Oct 31, 2025)
StatusOpen, Classified
ReasonUltrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining.
Root causeSoftware design
ActionOn November 1, 2025, Philips issues an Urgent Medical Device Recall (Correction) Notification to affected consignees. Phillips ask consignees to take the following actions: 1. Utilize an Android based device with your Lumify system. 2. Lightning port devices are not affected and may continue to be used as normal. 3. If you use an 05 based device, do not update your OS device to software version lax. 4. Turn off 05 auto-updates on USB host devices for Lumify by following the below steps: 5. Place this Notice with the documentation of the affected Ultrasound system(s); ensure the notice is in
Quantity4,819 systems
DistributionWorldwide US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR and the countries of Argentina, Arm
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203406Lumify Diagnostic Ultrasound System
Philips Ultrasound, Inc., Feb 23, 2021
K223771Lumify Diagnostic Ultrasound System
Philips Ultrasound, May 4, 2023
K232500Lumify Diagnostic Ultrasound System
Philips Ultrasound, Oct 26, 2023
K242519Lumify Diagnostic Ultrasound System
Philips Ultrasound, May 15, 2025
K242670Lumify Diagnostic Ultrasound System
Philips Ultrasound, May 8, 2025