Recall Z-0439-2026 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

B-K Medical A/S

bk3500 Ultrasound System w/battery; Model No. 2300-41; System, Imaging, Pulsed Doppler, Ultrasonic

DateNov 6, 2025 (initiated Oct 9, 2025)
StatusOpen, Classified
ReasonGE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
Root causeDevice Design
ActionGE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION NOTICE" to its consignees on 10/09/2025 via traceable means. The notice explained the issue, potential risk, and requested: "Actions to be taken by Customer / User You can continue to use your device by following the instructions below: 1. Review and follow the applicable addendums to the Instructions for Use and the updated Service Manuals. The Addendums and Service Manuals are available for download at: https://www.gehealthcare.com/support/manuals. 2. Follow the conditions defined within the addendum and Service Manual applicab
Quantity124 units
DistributionWorldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Estonia, Finland, France, Georgia, Germany, Greece, Hong
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223830Ultrasound System 2300
Bk Medical Aps, Apr 11, 2023