Recall Z-0435-2026 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

B-K Medical A/S

1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex Focus 300, Flex Focus 400, Flex Focus 500, Flex Focus 700, Flex Focus 800; System, Imaging, Pulsed Doppler, Ultrasonic

DateNov 6, 2025 (initiated Oct 9, 2025)
StatusOpen, Classified
ReasonGE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
Root causeDevice Design
ActionGE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION NOTICE" to its consignees on 10/09/2025 via traceable means. The notice explained the issue, potential risk, and requested: "Actions to be taken by Customer / User You can continue to use your device by following the instructions below: 1. Review and follow the applicable addendums to the Instructions for Use and the updated Service Manuals. The Addendums and Service Manuals are available for download at: https://www.gehealthcare.com/support/manuals. 2. Follow the conditions defined within the addendum and Service Manual applicab
Quantity4426 units
DistributionWorldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Estonia, Finland, France, Georgia, Germany, Greece, Hong
FDA recordOpen on fda.gov