Recall Z-0509-2026 · QNR Blood Pump For Ecmo, Long-Term (> 6 Hours) Use

Medtronic Perfusion Systems

MC3 VitalFlow Console, REF 58100; Blood pump of ecmo

DateNov 14, 2025 (initiated Oct 3, 2025)
StatusOpen, Classified
ReasonAs of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
Root causeSoftware in the Use Environment
ActionMedtronic Cardiac Surgery issued an Urgent: Medical Device Correction notice to its consignees on 10/03/2025. The notice explained the issue, potential harms, and requested the consignee contact Medtronic Sales Representative to schedule the service event to their affected VitalFlow" Console. Please reference Technical Service Update (TSU) "MOD0209" when scheduling. The notice also indicated that the consignees use unaffected VitalFlow Console devices, if available, until the correction has been completed. For questions regarding the communication, please contact your Medtronic Sales Rep
Quantity167 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K230364VitalFlow™ Console
Michigan Critical Care Consultants, Inc. (D.B.A Mc3 Inc.), Aug 25, 2023
K250199VitalFlow Console
Medtronic, Inc., May 20, 2025