Recall Z-0489-2026 ยท GKZ Counter, Differential Cell

Abbott Laboratories

Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS, 5.7 and 5.10.1 in the US Product Description: Alinity hq Analyzer Component: No

DateNov 12, 2025 (initiated Sep 18, 2025)
StatusOpen, Classified
ReasonSoftware issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.
Root causeSoftware design
ActionOn September 18, 2025, Abbott Laboratories issued an "Urgent Medical Device Recall" Notification. Abbott ask consignees to take the following actions: 1. If IG/%IG is above your laboratory established reference range (e.g. greater than 2%), review the results per your laboratory procedures. A manual smear review may need to be performed to confirm the results. 2. Complete and return the Customer Reply Form. 3. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. 4. Please retain this let
Quantity833
DistributionWorldwide distribution: US (nationwide) to states of: AL, CA, CO, FL, IN, MN, NC, NY, OK and OUS (foreign) to countries of: ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220031Alinity h-series System
Abbott Laboratories, Aug 4, 2023
K243283Alinity h-series System
Abbott Laboratories, Feb 20, 2025