Recall Z-0569-2026 ยท NJR Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test

Meridian Bioscience Inc

Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay

DateNov 19, 2025 (initiated Oct 7, 2025)
StatusOpen, Classified
ReasonThe cap on the reaction buffer tube may be missing the black rubber gasket resulting in the potential for leakage and reduced buffer volume.
Root causeNonconforming Material/Component
ActionMeridian Bioscience issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees beginning on 10/15/2025 via email. The notice explained the problem with the device, health hazard, and requested the following: "Required Actions: " Identify remaining inventory of Alethia C. difficile, Lot 480050U036. " Identify the Reaction Buffer bag located within the kit box. " Remove Reaction Buffer Tubes from Bag and inspect each cap for the presence of a gasket. " Segregate Reaction Buffer Tubes which have no gasket present. Record the quantity of tubes with no gasket present on the CONFIR
Quantity109 in total
DistributionUS Distribution to states of: AL, CO, FL, GA, IL, MD, ME, MT, NH, NJ, OH, TN, TX and WA; and OUS (foreign) countries of: Canada, Italy, and Panama.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121044ILLUMIGENE GROUP B STREPTOCOCCUS CARROT BROTH CLAIM
Meridian Bioscience, Inc., May 1, 2012