Recall Z-0569-2026 ยท NJR Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test
Meridian Bioscience Inc
Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay
| Date | Nov 19, 2025 (initiated Oct 7, 2025) |
| Status | Open, Classified |
| Reason | The cap on the reaction buffer tube may be missing the black rubber gasket resulting in the potential for leakage and reduced buffer volume. |
| Root cause | Nonconforming Material/Component |
| Action | Meridian Bioscience issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees beginning on 10/15/2025 via email. The notice explained the problem with the device, health hazard, and requested the following: "Required Actions: " Identify remaining inventory of Alethia C. difficile, Lot 480050U036. " Identify the Reaction Buffer bag located within the kit box. " Remove Reaction Buffer Tubes from Bag and inspect each cap for the presence of a gasket. " Segregate Reaction Buffer Tubes which have no gasket present. Record the quantity of tubes with no gasket present on the CONFIR |
| Quantity | 109 in total |
| Distribution | US Distribution to states of: AL, CO, FL, GA, IL, MD, ME, MT, NH, NJ, OH, TN, TX and WA; and OUS (foreign) countries of: Canada, Italy, and Panama. |
| FDA record | Open on fda.gov |