Recall Z-0453-2026 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare Finland Oy

Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/26 to include: CARESCAPE B450 MBA313 2095800-001 CARESCAPE MONITOR B450 (ATO MODEL) 2068491-002 CARESCAPE MONITOR B450 B1 2107633-001 CARESCAPE MONITOR B450-GS (V2.0.8) 2094080-00

DateNov 7, 2025 (initiated Sep 26, 2025)
StatusOpen, Classified
ReasonPotential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
Root causeComponent change control
ActionOn September 26, 2025, the recalling firm began notifying affected customers. A second letter was issued on October 3, 2025 which addressed a correction to Affected Product Details information. The customer letter advised users to use a monitor with a battery that is not within the affected battery population. However, pending correction by GE HealthCare, users may continue to use a monitor with affected batteries by following instructions included in the letter. These instructions include having mains power connected whenever possible, confirming appropriate battery status prior to use, and
Quantity3271
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213363CARESCAPE B450, E-musb
GE Healthcare Finland Oy, Apr 13, 2022