Recall Z-0543-2026 ยท QNW Reprocessed Bronchoscope

Auris Health, Inc

MONARCH Bronchoscope. Model Number: MBR-000211-B

DateNov 19, 2025 (initiated Oct 18, 2025)
StatusOpen, Classified
ReasonPotential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible to leaks.
Root causeProcess control
ActionAuris Health notified consignees via emailed "URGENT FIELD SAFETY NOTICE - DEVICE RECALL" letter dated 10/17/2025 on about 10/18/2025. Consignees were instructed to cease use of affected units immediately, forward the notification to any applicable personnel as well as customers if further distributed/transferred, and arrange for the return of all affected units. Consignees were also instructed to complete and return the Acknowledgement Form via email to MonarchSupport@its.jnj.com. Please return this form even if you no longer have or use the device subject to this notification. Please retur
Quantity1,477 units
DistributionNationwide distribution to AL, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KY, MA, MI, MS, MT, NC, NE, NJ, NM, NY, OH, OR, PA, SC, SD, TN, VA, VT, WA, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K231473Monarch Bronchoscope
Auris Health, Inc., A Johnson and Johnson Family Company, Nov 28, 2023